einstein (São Paulo). 01/Jan/2011;9(1 Pt 1):66-9.

Sterilization of single-use helical stone baskets: an experimental study

Fernando , Alex , Cely Barreto da , Roni de Carvalho , Marjo Deninson Cardenuto

DOI: 10.1590/S1679-45082011AO1762

ABSTRACT

Objectives:

To experimentally evaluate the efficacy of a standard sterilization protocol employed during reuse of disposable helical stone baskets.

Methods:

Study performed on 20 helical stone baskets: 10 were used in the initial validation process, contaminated with Escherichia coli ATCC 25922 and imprinted on Müeller-Hinton media; 10 catheters were contaminated with Geobacillus stearothermophilus ATCC 7953, processed, inoculated in TSB and incubated in a water bath at a temperature of 55°C. Bacterial growth was evaluated after 1, 3, 5 and 7 days. After sterilization, stone baskets were also opened and closed 40 times to check for functional problems. All plastic and basket parts were carefully checked for damages.

Results:

After the 72-hour incubation period, there was growth of E. coli ATCC 25922 in 100% of imprints. After the sterilization process and up to 7 days incubation period on a blood agar plate, there was no growth of G. stearothermophilus ATCC 7953 or any other bacteria. There were no functional problems or damage to baskets after the sterilization process.

Conclusion:

The ethylene oxide system is efficacious and safe for sterilization of disposable helical stone baskets. However, further clinical studies are required and should provide more safety information.